Aim: To investigate any relationship between sodium and potassium levels in blood gas analysis and the levels measured by otoanalyzer. Materials and methods: In our study, 2354 patient's test results of Na and K measurements simultaneously in blood gas and autoanalyzer were evaluated retrospectively for about 6 months period. In the emergency department Findings: Mean K values which is measured on blood gas and autoanalysers were 3.77 ± 0.84 mmol/l and 4.46 ± 0.80 mmol/l respectively. The mean difference was 0.68 ± 0.61 mmol/l for K. Mean Na values which is measured on blood gas and autoanalysers were 145.14±7.40 and 135.88±5.29 mmol/l respectively. The mean difference was -9.26±6.54 mmol/l for Na. Conclusion: It is not appropriate to use results of K and Na measurements in blood gas analysis instead of the results measured by the autoanalyzer
Keywords: Autoanalyzer, Blood gas analysis, Electrolyte.Amaç: Kan gazı analizi sırasında ölçülen Sodyum (Na) ve potasyum (K) un otoanalizörle elde edilen değerlerle eşit olup olmadığını araştırmak. Gereç ve Yöntem: Çalışmamızda 6 aylık sürede acil serviste eş zamanlı olarak kan gazı ve otoanalizörde Na ve K+ ölçümü yapılan 2354 hastanın tetkik sonuçları geriye dönük değerlendirildi. Bulgular: Kan gazında ve otoanalizörde ölçülen ortalama K+ değerlerini sırasıyla 3.77±0.84 mmol/l ve 4.46±0.80 mmol/l; ve K+ için fark ortalamasını 0.68±0.61 mmol/l saptadık. Kan gazında ve otoanalizörde ölçülen ortalama Na değerlerini ise sırasıyla 145.14±7.40 ve 135.88±5.29 mmol/l; ve Na için fark ortalamasını -9.26±6.54 mmol/l saptadık. Sonuç: Kan gazı analizi ile yapılan K+ ve Na ölçüm sonuçlarının otoanalizör ile yapılan ölçüm sonuçlarının yerine kullanılması uygun değildir.
Anahtar Kelimeler: Elektrolit, Kan gazı analizi, Otoanalizör.